Ysebaert · Model HS Compact
Allows user to archive more than 1 million results with patient identifications. By comparing the results to the preset reference values. Provides a first diagnostic of the patient’s pathology A smart system of solenoid valves combined with a unique automatic cleaning system makes this virtually maintenance-free. No more clogging or too frequent contaminations Large colour display Easy-to-use and really user-friendly. Allows the analysis of 8 different species, with preset normal ranges. Direct Dilution in the counting chambers Monosyringe Air / Liquid Monochrome LCD display
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump tubing Consumable - replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →