Beckman Coulter · Model Ac·T 5diff CP
Closed-tube sampling eliminates operators' exposure to biohazards Two sample tube holders accommodate tubes of varying sizes, such as pediatric tubes, for increased flexibility 60 sample-per-hour throughput enables a streamlined workflow Comprehensive onboard support database uses keyword searches to help users solve problems quickly and easily Specially formulated reagents, controls and calibrators ensure accuracy One-step setup, operation and shutdown speeds workflow Zero daily maintenance saves time and labor Automated quality control evaluation reduces manual errors Patient record storage enables streamlined data recall Qualitative flagging speeds review of patient results and decreases TAT 20 user-definable patient and action ranges increase flexibility Random access testing provides complete workflow flexibility CBC or CBC/Diff mode and patient range can be easily selected from the front keyboard display for easy programming.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, aperture-clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn segment / shear valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Tubes not advancing, cassette / carousel error, mixing-station fault.
Root cause: Worn autoloader belt / gripper, a mis-loaded rack, or a mixing-station fault.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Segment / shear valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| VCS flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
| Autoloader belt / gripper (DxH) | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The recall was initiated because Beckman Coulter has confirmed that an issue has been identified with one of the raw materials used to manufacture the Coulter A T 5 diff PN 8547171.
The recall was initiated because Beckman Coulter identified a problem with occasional unexpectedly high results for RBC, HCT, PLT and MPV without instrument generated messages, for the first run of a capped patient sample. Upon repeat analysis the results are correct. Patients could be affected as described below: (1) False high RBC and HCT may result in a delay of treatment, or the delivery of
Software anomaly causes sample results to be not credible, reported as '0' or left blank. The next sample will also be incorrectly reported for the same parameter.
Samples with high fat content will not count platelets accurately.