Extracted directly from this unit's own service manual.
1000Out of memory in %s Fatal system or program error!
Check the COMB card and Dimm-PC module for proper insertion. Try to make a software upgrade.
1001Error %d on opening file %s Fatal system or program error!
The same as in case of error code 1000.
1002Cannot create file %s Fatal system or program error!
The same as in case of error code 1000.
1003Data file %s is corrupt, new will be created, all stored measure data will be lost
Stored data file is corrupt or missing! Disk or software error. New stored data file will be created automatically (by a User confirmation).
1004Error %d on indexing file %s Fatal system or program error!
The same as in case of error code 1000.
1100Pressure error. Cannot make the measuring vacuum!!
Check the pump and the tubes, fittings and valves around the puffer reservoir.
1101Waste is full!
Empty waste container! The system assumes that you will do it before going on!
1102Check Diluent container! Diluent tank is empty, or the dil./clean sensor is too sensitive (try recalibrating the sensors), or there are lot of bubbles in the diluent tube
Replace the diluent reagent container with a filled one. If this is still a problem with a filled diluent container, run Reagent sensor calibration and check the tubing and fittings of diluent path.
1103Check Lyse container! Lyse tank is empty, or the lyse sensor is too sensitive (try recalibrating the sensors), or there are lot of bubbles in the lyse tube
Replace the lyse reagent container with a filled one. If this is still a problem with a filled lyse reagent container, run Reagent sensor calibration and check the tubing and fittings of lyse path.
1105Vacuum error The measuring vacuum has been dramatically decreasing during the measurement or absolutely lost!
Leakage in the pneumatic system (check the pump, fittings, tubing, pressure sensor, puffer reservoir and valves).
1106Pneumatical error occurred! Device: %s Error code: %s Status: %x Retry?
Fatal error in mechanical subsystem. If frequently happens this error check the proper operation of the motor modules (home/end opto switches, ribbon cables and moving of motors).
1107Check Cleaner container! Cleaner tank is empty, or the dil./cleaner sensor is too sensitive (try recalibrating the sensors), or there are lot of bubbles in the cleaner tube
Replace the cleaner reagent container with a filled one. If this is still a problem with a filled cleaner container, run Reagent sensor calibration and check the tubing and fittings of cleaner path.
1110Check Rinse container! Rinse tank is empty, or the rinse sensor is too sensitive (try recalibrating the sensors), or there are lot of bubbles in the rinse tube
Replace the rinse reagent container with a filled one. If this is still a problem with a filled rinse container, run Reagent sensor calibration and check the tubing and fittings of rinse path.
1111The sampler rotor was jammed during sampling process! Retry sampling?
Cap was not taken off from the sample tube, or tube is not placed correctly, or other mechanical jam. Check the adapter and the tube. Press Retry or Stop.
1112The sampler rotor was jammed! Mechanical problems like adapter is moved out from the right place
Check the adapter and the tube. Press the button
1113Unrecoverable pneumatical error occurred! Device: %s Error code: %s Status: %x Please turn off the instrument and turn it on again!
Fatal error in mechanical subsystem. Possibly mechanical jam. Check the proper operation of the motor modules (home/end opto switches, ribbon cables and the motor moving).
1200Fatal MeasInit error.
Power off the system! Fatal COMB error. Replace the COMB board.
1201Fatal HGB error. HGB channel did not give a ready signal!
Power off the system! Replace the COMB board.
1300Cannot write archive data! Check disk! It may be unformatted, full or bad! Do you want to retry?
Cannot write archive data because some disk errors! Change the floppy disk.
1301Cannot read archive data! Do you want to retry?
Archive disk is damaged or not an archive disk! Change the floppy disk (archive).
1302Cannot load archive data. Disk is not an archive disk or damaged!
Change the floppy disk (archive).
1400Printer error %d Printer error! Check the connections between instrument and printer and check printer setup in Printer Settings
Check the connections between instrument and printer and check printer setup in Printer Settings
1401Printer is out of paper and error %d Printer is out of paper and there are some other printer errors!
Feed the printer with paper. Check the connections between the instrument and the printer and check the printer setup in the Settings/Printer settings sub-menu.
1402Printer is out of paper Printer is out of paper!
Feed the printer with paper.
1403Printer was not set up correctly The selected printer type does not match to the printer!
Modify the printer setup in the Settings/Printer settings sub-menu.
1404Paper width is too large Paper width overstep the margin.
Modify the paper setup in the Settings/Printer settings sub-menu.
1405Paper height is too large Paper height overstep the margin!
Modify the paper setup in the Settings/Printer settings sub-menu.
1406Paper width is too small Paper width overstep the margin!
Modify the paper setup in the Settings/Printer settings sub-menu.
1407Paper height is too small Paper height overstep the margin!
Modify the paper setup in the Settings/Printer settings sub-menu.
1408Left margin is too large Left margin is too large!
Modify the margin setup in the Settings/Printer settings sub-menu.
1409Top margin is too large Top margin is too large!
Modify the margin setup in the Settings/Printer settings sub-menu.
1410Spacing is too large The distance between two results is too high!
Modify the vertical spacing setup in the Settings/Printer settings submenu.
1411Error #%d in printer initialization! Printer software error!
Check the connections between the instrument and the printer and check the printer setup in the Settings/Printer settings sub-menu. Try to make a software upgrade.
1900You have to add at least one user to use multi user mode!
Multi-user problem. Follow the instruction, or change to single user mode in the User settings submenu.
1921%s sensor calibration failed! The system turned off the %s sensor.
Reagent container became empty, or there are lot of bubbles in reagent tubes. Replace the reagent container with a filled one. If this is still a problem with a filled reagent container, turn on the sensor and try recalibrating and check the tubings and the fittings.
5001Remove reagent tubing at rear reagent inputs (Diluent, Lyse, and Cleaner).
5002Connect min. 100 ml distilled water to reagent inputs using cleaning tube kit.
5003Remove cleaning tube kit. Keep reagent inputs free.
5004You can power off the system!
5005Apertures are partially clogged. Try cleaning!
5006Please empty waste container!
5100%u data record(s) will be copied to the database.
5101%u data record(s) is selected.
5102Insert data disk!
5103Insert data disk #1 of 1!
5104xx data record(s) will be saved on 1 disk(s). Insert an empty floppy disk!
5105Insert next empty floppy disk!
5106No such type of data
5200Your previous QC data will be lost. Are you sure?
5300Some or all of the calibration factors are out of range!
5400+12V too low!
5401+12V too high!
5402Battery voltage too low!
5900This function is not available because there are no accepted blank values. Try blank measurement!
5901Insert software update disk! The system will restart.
5902Are you sure to delete all device statistics?
5903This function will delete all measured results. Are you sure?
5904This function will delete all selected records. Are you sure?
5905Are you sure?
5906You can power off the system now or restart it by pressing any key!