Sysmex · Model XS-1000i
Effective diagnostic capabilities utilizing fluorescent flow cytometry CBC and 5-part differential Simplicity in operation for streamlined workflow Standardization across Sysmex platform Proven reliability as a member of the X-Family of products For better reliability, longer life and more efficiency the XS-1000i utilizes an avalanche photodiode with a small laser bench and integrated electronic boards. For enhanced diagnostics, WBC cell populations are measured using fluorescent flow cytometry with side scatter and side fluorescence.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
Consumable, replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
| Peristaltic pump tubing Consumable, replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →