Beckman Coulter · Model LH 750
Performance and Results that Exceed Your Expectations The LH 750 analyzer is equipped with superior technologies, NRBC enumeration and random access capabilities. These advancements help high-volume labs eliminate manual pre-sorting of samples and speed results to clinicians. With this system, you can improve overall efficiency, while maintaining the highest degree of testing accuracy.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, aperture-clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture.
Symptoms: Tubes not advancing, cassette / carousel error, mixing-station fault.
Root cause: Worn autoloader belt / gripper, a mis-loaded rack, or a mixing-station fault.
Symptoms: Short samples, aspiration errors, cap-piercing failures.
Root cause: Worn or bent sample probe / closed-vial piercer.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms; note the LYSE-lot recall history.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Haemoglobin or reticulocyte results biased, blank absorbance off, optics alarm.
Root cause: Photometer lamp aging, a dirty flow cell, or the retic optics.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Autoloader belt / gripper | Periodic replacement | $$ | Moderate |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| HGB photometer lamp / VCS flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Beckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or reticulocytes.
This recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS if the year is greater than 2010. The following dates are affected: Run date, Control expiration date, Date of birth (For individuals born in 2011 or later), a
A rare risk of sample misidentification exists with Coulter LH 750 Hematology Analyzers. The printout results from the instrument are affected; however, the results stored in the workstation's database are not affected.