Nihon Kohden · Model Celltac Alpha MEK-6500K
Built-in open and closed tube mode To reduce the risk of contamination, the Celltac α MEK-6500K includes both, an open and a closed tube mode for easy blood sampling. Auto dilution mode change You can set panic value thresholds (abnormal high and low values) to trigger re-measurement in preset dilution ratio modes (low, normal, high, higher). In higher mode the measuring range for WBC can be extended to 599 x 10³/µl while the low dilution mode gives high accuracy even in low values of WBC or PLT. Capillary mode The Celltac α MEK-6500K allows you to analyze capillary blood with only 10 µl. This is the ideal method to perform CBCs inclusive 3-part-diff for pediatric and geriatric patients. Unlimited patient memory The instrument can store unlimited patient samples together with QC results and alarm logs by using SD-card memory (2GB can store 30 000 patient samples).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.