Nihon Kohden · Model Cap-STAT OLG-2800
Large display for waveform and numeric display CO2 and Respiration Alarms Compact and lightweight AC and battery operation Wireless connection to Prefense™ Early Detection and Notification System™ Post anesthesia care Sedation for pain control Conscious sedation procedures Wireless connection to Prefense™ Early Detection and Notification System™ Uses Cap-ONE CO2 sensor system.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: EtCO2 offset, slow response, or a zero-required prompt.
Root cause: Sample cell / airway adapter contamination or moisture, or IR bench drift.
Symptoms: Flat trace, occlusion or "check adapter" message.
Root cause: Blocked / kinked sample line, a full water trap, or a sample-pump fault.
Symptoms: Host monitor does not display the capnography.
Root cause: Docking / connector or cable fault.
Symptoms: Repeated occlusion or contamination flags.
Root cause: Reused single-patient line or trap, or wrong type.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water trap / sampling line | Consumable | $ | Easy |
| CO2 module / bench service | Periodic replacement | $$$ | Fairly easy |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
| Airway adapter (mainstream) | Consumable | $ | Easy |
| Module docking connector | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.