Masimo · Model EMMA Capnograph
Clear, continuous capnograph of carbon dioxide values Simple, easy-to-use interface for quick set-up and one-touch programming Audible and visual alarm system for No Adapter, Clogged Adapter, No Breath (Apnea), Low Battery and adjustable High and Low EtCO2 alarm.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No power on, startup sequence does not finish.
Root cause: No battery fitted or depleted batteries.
Symptoms: Measured EtCO2 outside specified accuracy, drift versus reference gas.
Root cause: Incorrect zero reference - a zeroing procedure is required, then verify with reference gas.
Symptoms: Persistent "No Adapter" or "Clogged Adapter" alarm, apnea alarm with a breathing patient.
Root cause: Adapter not seated, a used/contaminated single-use adapter, or moisture/secretions in the adapter window.
Symptoms: Numbers appear dim.
Root cause: Automatic brightness control activated by bright light or sunlight - pressing any button restores normal brightness.
Symptoms: No data on the paired device, connection drops.
Root cause: Bluetooth off, out of range, wrong pairing, incompatible or missing app, or damaged Bluetooth hardware.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Replace immediately when the battery indicator blinks; carry spares.
Single-patient disposable; do not reuse.
Verification tool, not a spare part.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| AAA batteries (x2) Replace immediately when the battery indicator blinks; carry spares. | Consumable | $ | Easy |
| EMMA airway adapter Single-patient disposable; do not reuse. | Consumable | $ | Easy |
| Display / housing assembly | Major service part | $$$ | Difficult |
| Reference CO2 gas / verification kit Verification tool, not a spare part. | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.