CareFusion · Model Smoke check
Results are displayed instantly on a clear graphic display using colour-coded bands indicating non smoker, light smoker and heavy smoker. Results can be converted to %COHb. No warm up required. Clear "traffic light" indication of CO levels. Push button calibration procedure. Auto-zero function to allow immediate tests of numerous patients Small, portable and lightweight with a textured handgrip Supplied complete with all accessories in a sturdy carry case.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Reading does not match a reference gas, colour bands / %COHb wrong, drift over time.
Root cause: Electrochemical CO sensor aged or poisoned, or the push-button calibration not performed / failed.
Symptoms: Unit will not zero between patients, baseline offset, cannot start a test.
Root cause: Sensor drift or residual CO in the sample path; sensor at end of life.
Symptoms: Missing segments, wrong colour band, dim display.
Root cause: Display or control-PCB fault.
Symptoms: Slow or no response, moisture in the path, cross-patient concern.
Root cause: Contaminated or wet single-use mouthpiece / D-piece, or a blocked sample path.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Use only Micro Medical / CareFusion parts; calibrate after fitting.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electrochemical CO sensor Use only Micro Medical / CareFusion parts; calibrate after fitting. | Periodic replacement | $$ | Moderate |
| Single-use mouthpiece / D-piece | Consumable | $ | Easy |
| Calibration gas / adapter | Periodic replacement | $$ | Easy |
| Battery | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.