Masimo · Model ISA OR+
Minimal warm-up time and readings in seconds Monitoring patients with high respiration rates, up to 150 bpm Automatic temperature and pressure compensation Low power consumption Provides minimal alveolar concentration (MAC) calculated from the measured anesthetic agents and N2O* Appropriate for monitoring of infant, pediatric, or adult patients in a range of clinical environments including the OR, ICU, Medical-Surgical units, and emergency Compatible with Masimo’s Nomoline™ Airway Adapter Set.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: EtCO2 offset, slow response, or a zero-required prompt.
Root cause: Sample cell / airway adapter contamination, or IR bench drift.
Symptoms: Flat trace, occlusion or "check adapter" message.
Root cause: Blocked / kinked sample line, a full Nomoline water trap, or a sample-pump fault.
Symptoms: Host monitor does not display the ISA module.
Root cause: Docking / cable connector fault.
Symptoms: Repeated occlusion or contamination flags.
Root cause: Reused single-patient adapter / Nomoline, or wrong adapter type.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Nomoline sampling line / water trap | Consumable | $ | Easy |
| ISA analyser module | Major service part | $$$ | Fairly easy |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
| Mainstream airway adapter | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.