DRE · Model Echo CO2
5.7 inch high resolution display for easy reading Lightweight, portable design and user-friendly interface for easy operation Flexible configurations to meet different clinical needs PR measurement Respironics Loflo sidestream or CAPNOSTAT 5 EtCO2 mainstream measurement Powerful storage capacity Real-time parameters measurement display with trend table for easy reviewing Built-in rechargeable Lithium-ion battery for 10 hours continuous working Bi-directional communications with central station by wired or wireless network.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: EtCO2 does not match a reference, slow response, baseline offset.
Root cause: CO2 sensor aging, or the user CO2 calibration not performed.
Symptoms: Flat trace, occlusion or "check adapter" message.
Root cause: Blocked / wet airway adapter or sample line, or a sensor-cable fault.
Symptoms: No saturation value with a good site.
Root cause: SpO2 sensor or cable fault.
Symptoms: Short run time, will not power on.
Root cause: Depleted or aged battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CO2 sensor / bench | Periodic replacement | $$$ | Moderate |
| Airway adapters | Consumable | $ | Easy |
| Sample line / water trap | Consumable | $ | Easy |
| SpO2 sensor and cable | Frequent wear item | $$ | Easy |
| Battery | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.