NMNova Metrix · Model Tidal Wave
Y-site probe included Offering accurate, reliable monitoring in a convenient, portable package.deal for use in emergency and transport settings, and other alternate site monitoring situations. Hand-held mainstream & sidestream capnograph 30 minute on-screen graphic trend available for assessing outcome of therapeutic intervention Visual and Audible "No Respiration" alert provides for a higher level of patient safety during transport and other alternate site monitoring situations Display is visible in any lighting condition - from total darkness to bright sunlight User replaceable disposable and rechargeable battery packs eliminate the need to wait while batteries are charging. Easy user CO2 calibration Available with SpO2.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: EtCO2 does not match a reference, slow response, baseline offset.
Root cause: CO2 sensor aging, or the user CO2 calibration not performed.
Symptoms: Flat trace, "no breath" alert with the patient breathing.
Root cause: Fouled or wet mainstream airway adapter / sidestream line, or a sensor cable fault.
Symptoms: Unit powers off quickly, will not hold charge.
Root cause: Aged rechargeable pack; disposable AA pack depleted.
Symptoms: No saturation value with a good sensor site.
Root cause: SpO2 sensor or cable fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CO2 sensor / bench | Periodic replacement | $$$ | Moderate |
| Mainstream airway adapters | Consumable | $ | Easy |
| Rechargeable battery pack | Periodic replacement | $$ | Easy |
| SpO2 sensor and cable | Frequent wear item | $$ | Easy |
| Sidestream sample line / water trap | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio