Nomoline™–No moisture sampling lines and non-proprietary cannulas Reduces disposable costs through: Extended monitoring time in low- and high-humidity environments Use of non-proprietary cannulas Innovative design eliminates the need for a water trap Patented design allows water in the sampling line to evaporate into the surrounding air, while leaving oxygen, carbon dioxide, and anesthetic gases unaffected Hydrophobic liquid and bacteria filter protects the ISA module from bacteria and water intrusion to maintain continuous performance.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: EtCO2 offset, slow response, or a zero-required prompt.
Root cause: Sample cell / airway adapter contamination, or IR bench drift.
Symptoms: Flat trace, occlusion or "check adapter" message.
Root cause: Blocked / kinked sample line, a full Nomoline water trap, or a sample-pump fault.
Symptoms: Host monitor does not display the ISA module.
Root cause: Docking / cable connector fault.
Symptoms: Repeated occlusion or contamination flags.
Root cause: Reused single-patient adapter / Nomoline, or wrong adapter type.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Nomoline sampling line / water trap | Consumable | $ | Easy |
| ISA analyser module | Major service part | $$$ | Fairly easy |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
| Mainstream airway adapter | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.