Mindray · Model BC-5300
Semi-conductor laser scatter to differentiate white blood cells by cell size and granule complexity Patented cluster algorithm to separate cell groups with great precision Intuitive flagging information to help determine abnormal samples Supports both whole blood and capillary blood samples in open vial 20uL blood sampling for CBC+DIFF test and 15uL for CBC test Up to 60 samples per hour throughput Economical reagent consumption Customization on reference range, report styles and auto-cleaning schedule Day-to-Day QC monitoring Bi-directional LIS connection
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump tubing Consumable - replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.