Mindray · Model BA-88A
7.0 TFT touch-screen & pop-up keypad. Two test mode: flowcell or cuvette. Supports bi-chromatic tests for end point, fixed-time, kinetics methods. External keyboard or mouse support via USB. User-friendly interface and multi-language operation software. Built-in thermal printer.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blank or standard absorbance off, poor precision, QC out of range.
Root cause: Lamp aging, a dirty flow cell / cuvette, or a filter-wheel fault.
Symptoms: Short samples, bubbles in the flow cell, no aspiration, clot alarm.
Root cause: Worn peristaltic pump tubing, a blocked flow cell, or a probe / valve fault.
Symptoms: Flow-cell / cuvette temperature not at 37 C, temperature-sensitive results drift.
Root cause: Heater or temperature-sensor fault.
Symptoms: No printout, unresponsive keypad.
Root cause: Printer or display fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp | Periodic replacement | $$ | Moderate |
| Peristaltic pump tubing | Frequent wear item | $ | Fairly easy |
| Flow cell / cuvette | Periodic replacement | $$ | Moderate |
| Filter wheel | Major service part | $$ | Moderate |
| Heater / temperature sensor | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio