Extracted directly from this unit's own service manual.
1In ascending action of the stirrer, it does not reach the upper dead point (on the rinsing bath side). (Alarm at the first upper dead point after resetting will be issued from other than the cell side.)
Check the upper dead point detector.
1 2In ascending action of the stirrer, it does not reach the upper dead point (on the cell side).
Same as above
1 3In descending action of the stirrer, it does not leave the upper dead point.
Same as above
1 4When the stirrer moves toward the rinsing bath, it does not reach the rinsing bath position.
Check the home detector.
1 5The stirrer does not come to the cell position.
Check the detector on the cell side.
1 6At resetting, the stirrer does not reach the rinsing bath position (home position) in its return movement to the home position.
Check the home detector.
1 7At resetting, the stirrer does not leave the rinsing bath position (home position) in its departing movement from the home position.
Same as above
1 8In rotation of the stirrer, it is not set at the upper dead point.
Check the upper dead point detector.
3The rinsing mechanism does not reach the upper dead point in ascending motion.
Check the upper dead point detector.
3 2The rinsing mechanism does not leave the upper dead point in descending motion.
Same as above
5The reaction disk cannot recognize its stop position.
Check the detectors for stop positions on the inner and outer tracks.
5 2The reaction disk does not stop at the specified position.
Same as above
5 3At resetting, the reaction disk cannot recognize its home position.
Check the home detector.
5 4At resetting, the first cell on reaction disk does not stop at the specified position.
Check the home detector or inner/outer track detector.
5 5When the reaction disk turns, the serum probe, reagent probe, stirrer or rinsing mechanism is not set at the upper dead point (on the cell side).
(1) Perform resetting. (2) Check the upper dead point detector of the mechanism which caused alarm.
61The serum probe does not reach the upper dead point in ascending motion (on other than the cell side). (Alarm at the first upper dead point after resetting will be issued from other than the cell side.)
Check the upper dead point detector.
62The serum probe does not reach the upper dead point in ascending motion (on the cell side).
Same as above
64The serum probe moves down abnormally in descending action (on the cell side).
Same as above
65The serum probe does not go down from the upper dead point in descending motion.
Check the upper dead point detector.
66The serum probe does not go down from the upper dead point in descending motion (on the cell side).
Check the upper dead point detector.
67Detection of abnormal descending motion of the serum probe remains on.
Refer to alarm code 6-3.
68When the serum probe turns to the cell side, the cell position cannot be detected.
Check the home detector.
69When the serum probe turns from the cell side to other position, it does not come off the cell position.
Same as above
6 11Before the probe goes down, the liquid level detector is already turned on.
Refer to alarm code 6-3.
6 12The serum probe moves down abnormally in descending action (only at turning of the serum probe in adjustment).
Same as above
6 13In rotation of the serum probe, it is not set at the upper dead point.
Check the upper dead point detector.
6 14In rotation of the serum probe from the reset position, it does not come off the reset position.
Check the home detector.
6 15In rotation of the serum probe to the reset position, it cannot detect the reset position.
Same as above
81The sample disk cannot detect the stop position on outer track.
Check the outer track detector.
82The sample disk does not stop at the specified position on outer track.
Same as above
83The sample disk cannot detect the stop position on inner track.
Check the inner track detector.
84The sample disk does not stop at the specified position on inner track.
Same as above
85At resetting, the sample disk cannot detect the home position.
Check the home detector.
86At resetting, the sample disk does not stop at the specified point of home position.
Same as above
87At resetting, the sample disk does not leave the home position.
Check the home detector.
11 1 to 60In sipping from the sample cup, sample on the sample disk is inadequate. Sub-code groups: Routine/stat samples: 1 to 35; Control STD: 36 to 57; W1 to W3: 58 to 60
Add sample.
15 1The serum syringe does not reach the upper dead point.
Check the upper dead point detector.
15 2The serum syringe does not go down from the upper dead point.
Same as above
16 1The reagent probe does not reach the upper dead point in ascending motion.
Check the upper dead point detector.
16 2The reagent probe moves down abnormally in descending motion.
Check liquid level detection.
16 3The reagent probe does not go down from the upper dead point in descending motion.
Check the upper dead point detector.
16 4Detection of abnormal descending motion of the reagent probe remains on.
Check the descending error detector.
16 5When the reagent probe turns toward the cell, it cannot detect the cell position.
Check the home detector.
16 6When the reagent probe turns from the cell side to other position, it does not leave the cell position.
Same as above
16 7Before the probe goes down, the liquid level detector is already turned on.
Check liquid level detection.
16 8In rotation of the reagent probe, it is not set at the upper dead point.
Check the upper dead point detector.
18 1The stop position of reagent disk cannot be detected.
Check home detection or count detector.
18 2The reagent disk does not stop at the specified position.
Same as above
18 3The home position of reagent disk cannot be detected.
Check home detection.
22 1The reagent syringe does not reach the upper dead point.
Check the upper dead point detector.
22 2The reagent syringe does not move down from the upper dead point.
Same as above
24 1The sipper nozzle does not reach the upper dead point (during resetting/operation).
Check the upper dead point detector of ISE sipper.
24 2The sipper nozzle does not leave the upper dead point.
Same as above
26 1The sipper syringe does not reach the upper dead point.
Check the upper dead point detector of ISE Syringe.
26 2The sipper syringe does not leave the upper dead point.
Same as above
27 1The ISE function is stopped due to alarm. (This warning is indicated when restart in the sampling stop status was attempted.)
Check alarm log and deal with the alarm.
28 1The water temperature of incubation bath is higher than 45.0°C.
Check the thermistor or thermostat of heater.
28 2The water temperature of incubation bath is outside a range of 37 ± 0.5°C.
Same as above
29 1The water level of incubation bath is too low.
Replenish water or check the drain solenoid valve.