Human Diagnostics · Model HumaStar 200
HumaStar 200 is the first choice for laboratories processing up to 650 photometric measurements/shift. Continuous sample loading and STAT processing at any time makes the HumaStar 200 a perfect match for small routine or emergency laboratories or for being used as dedicated test or backup analyzer in bigger ones. The extraordinary low water consumption of less than 2l/h grants autonomy from water supply units.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →