Horiba Medical · Model ABX e-Sat
Touch screen interface Printer Barcode reader Front USB port to archive data with easer User identification Patient identification by barcode or keyboard input Quality control management (results and time interval QC) Reagent management (expiration and volume) Optimized presentation of results to accommodate various clinical environments Search and display engine for 1000 archived results 1000 patient results Built-in search engine Exportable data (USB key) Parameters (16 + 2) CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV, *RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA Graphics: LMG histogram, RBC, PLT histograms Sample-Volume: CBC mode: 10 µL Stability: 24 hours (CBC) Operating sample mode: Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL) Identification: Patient I.D., Sequence# Flags: Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags RUO (Research Use Only)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.