HESKA · Model DRI-CHEM 7000
• Electrolytes and More: Electrolytes on one slide, individual slide options, and pre-packaged panels including the Comprehensive Plus, let you configure the profiles you require. • Completely Automated Dilution: No guesswork on measuring dilutions or calculating results. The DRI-CHEM® 7000 analyzer provides automated dilution and results calculations on every parameter for samples outside of measurable limits. • Sampling Flexibility:Choose the 1.5-ml heparinized sample tube or pipette serum or plasma from your regular red-top or green-top tubes into a non-heparinized tube. • Stack and Run: Load up to 22 tests (20 slides) at one time without jamming, using the convenient stackable slide rack. • Whole Blood Solution: Heska’s lithium heparin sample tubes save you a step. Go from syringe to centrifuge to analyzer with no pipetting required.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump tubing Consumable, replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →