GE HealthCare · Model Mac-Lab
Reliability - Combines Windows NT reliability and usability with standard off-the-shelf computer hardware for proven dependability. Productivity - The network capabilities of the Mac-Lab 7000 include workstations for remote real-time review as well as simultaneous data entry.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: IBP trace offset, "zero required" persists, pressures implausible.
Root cause: Transducer not open to atmosphere during zero, a damaged pressure cable, or an amplifier / acquisition fault.
Symptoms: Surface or IC electrograms noisy, missing channels.
Root cause: Patient cable / junction box, catheter connections, or amplifier noise / grounding.
Symptoms: Channels drop, the review workstation hangs, case will not start.
Root cause: Amplifier module or workstation PC fault.
Symptoms: No waveform to the lab displays or the recording system.
Root cause: Video / network output fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pressure cable set | Periodic replacement | $$ | Easy |
| Patient cable / junction box | Periodic replacement | $$$ | Easy |
| Acquisition amplifier module | Major service part | $$$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
| Video / network output hardware | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.