Philips Healthcare · Model Xper Flex Cardio Physiomonitoring System
Space-saving design delivers placement and positioning flexibility Streamlined FFR workflow makes it easy to access FFR measurements, from acquisition to automatic data entry onto reports DXL Algorithm provides 16-lead ECG analysis, Culprit Artery Detection and patented ST Maps.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: IBP trace offset, "zero required" persists, pressures implausible.
Root cause: Transducer not open to atmosphere during zero, a damaged pressure cable, or an amplifier / acquisition fault.
Symptoms: Surface or IC electrograms noisy, missing channels.
Root cause: Patient cable / junction box, catheter connections, or amplifier noise / grounding.
Symptoms: The Flex acquisition module drops out or will not connect to the review system.
Root cause: Module battery, the wireless / docked link, or a module fault.
Symptoms: No waveform to the lab displays or the recording system, review workstation hangs.
Root cause: Video / network output or workstation fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pressure cable set | Periodic replacement | $$ | Easy |
| Patient cable / junction box | Periodic replacement | $$$ | Easy |
| Flex acquisition module / battery | Periodic replacement | $$$ | Fairly easy |
| Workstation PC | Major service part | $$$ | Moderate |
| Video / network output hardware | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.