Lidco · Model Rapid
Early and fast warning of change. Clear indication of therapeutic route: fluid or drug. Quantification of hemodynamic response particularly stroke volume. More effective use of fluids the right amount at the right time. Contributes to reduced morbidity and complications particularly infections. Reduced length of stay. Reduced overall cost of care. Facilitates SV optimisation while displaying: > Pressures MAP, Systolic and Diastolic > Heart Rate > Stroke Volume and Cardiac Output (Scaled or Actual) > Dynamic Preload parameters Pulse Pressure Variation (PPV) and Stroke Volume Variation (SVV)8 > User selected event response window with continuous trending > History Screen with up to 24 hours of graphical trends > Chart screen for recording numeric data > User adjustable time scales.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: nominal SV / CO dashed or jumps, poor trace-quality indicator.
Root cause: Damped or poor arterial pressure waveform (air, clot, kinked line, under-damped tubing), or the pressure input not zeroed.
Symptoms: Pressure trace absent, noisy, or will not zero.
Root cause: Pressure interface cable, the connection to the patient monitor's analog output, or the transducer.
Symptoms: No response or offset touches.
Root cause: Touchscreen controller fault.
Symptoms: Shuts down off mains, battery alarm.
Root cause: Aged internal battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Arterial pressure interface cable | Periodic replacement | $$ | Easy |
| Internal battery | Periodic replacement | $$ | Fairly easy |
| Touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.