Fresenius · Model 4008S
The effective in-vivo urea clearance (K) The accumulated cleared plasma (Kt) or the current dialysis dose administered (Kt/V) The plasma sodium concentration: possible deviations to the prescribed therapy goal can be detected and corrected immediately during treatment* without incurring additional expenses for disposables or staff efforts. Advanced dialysis fluid circuit with highly-precise volumetrically controlled ultrafiltration Consistently reliable hydraulics. High-quality HD treatment to protect your patient. Easy, fast and intuitive handling TDMS (Therapy Data Management System) BPM (Blood Pressure Monitor) CDS (Central Concentrate Delivery System) AdaptedFlow1.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Conductivity out of window, bypass, treatment cannot start.
Root cause: Wrong or empty concentrate, air in the concentrate line, or a mixing-chamber / conductivity-cell fault.
Symptoms: "Blood leak" or "air" alarm, blood pump stops, venous clamp closes.
Root cause: Genuine dialysate discolouration, a dirty blood-leak-detector optic, air in the venous line, or a detector fault.
Symptoms: Arterial, venous or TMP pressure out of limits, frequent alarms.
Root cause: Kinked lines, a clotting dialyser, or a pressure-transducer / protector fault.
Symptoms: UF volume error, fluid-balance alarm, treatment halts.
Root cause: Balancing-chamber diaphragm or valve fault.
Symptoms: Temperature out of range, bypass.
Root cause: Heater or temperature-sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Balancing chamber diaphragm / valve kit | Periodic replacement | $$$ | Difficult |
| Conductivity cell | Periodic replacement | $$ | Moderate |
| Blood-leak detector assembly | Major service part | $$ | Moderate |
| Pressure-transducer protectors | Consumable | $ | Easy |
| Dialysate heater / temperature sensor | Major service part | $$ | Difficult |
| Blood pump rotor | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.