Fisher and Paykel Healthcare · Model MR850
One button selects optimal temperature and humidity levels for adult, pediatric and neonatal patients. Easy set up with simple controls and an easy to read display. Clinically based alarm system. Advanced algorithms and the ability to sense flow optimises humidification delivery and minimises nuisance alarms. Automatic standby during periods of gas flow interruption or water out.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Low airway temperature, condensation or dryness, temperature not reaching set point.
Root cause: Water chamber not seated on the heater plate, a failed heater base plate, or a temperature-probe fault.
Symptoms: Heater-wire fault alarm, circuit not heating, rain-out in the tube.
Root cause: Open heated breathing circuit, a faulty heater-wire adapter, or (MR730) the F2 fuse.
Symptoms: Implausible temperature, probe-fault alarm, unit will not run.
Root cause: Damaged airway or chamber temperature probe or its connector.
Symptoms: Unit cannot hold the set gradient, alarms and reverts to fail-safe.
Root cause: Control-board or sensor fault.
Symptoms: Chamber runs dry, water-out alarm, auto-feed not maintaining level.
Root cause: Empty supply bag, a stuck float valve, or a cracked chamber.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-patient or per-protocol consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater base plate | Major service part | $$$ | Difficult |
| Temperature probe set (airway + chamber) | Frequent wear item | $$ | Easy |
| Heater-wire adapter | Periodic replacement | $ | Fairly easy |
| Water chamber Single-patient or per-protocol consumable. | Frequent wear item | $ | Easy |
| Auto-feed / float valve assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.