Hudson RCI · Model ConchaTherm III
Hudson RCI ConchaTherm III 380-80 heated humidifier offers the utmost in simplicity, while providing the option to make adjustments and customize settings in response to unique patient or environmental challenges or in keeping with hospital protocols.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Dry gas to the patient, low airway temperature, chamber not warming.
Root cause: Heater base plate fault, chamber not seated, a temperature-probe fault, or a water-feed issue.
Symptoms: Rain-out in the tube, heater-wire fault alarm.
Root cause: Open heated breathing circuit or a heater-wire adapter fault.
Symptoms: Implausible temperature, probe-fault alarm, unit will not run.
Root cause: Damaged airway or chamber temperature probe or its connector.
Symptoms: Chamber runs dry or overfills, water-out alarm.
Root cause: Float valve or auto-feed fault, or a cracked chamber.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater base plate | Major service part | $$$ | Difficult |
| Temperature probe set (airway + chamber) | Frequent wear item | $$ | Easy |
| Heater-wire adapter | Periodic replacement | $ | Fairly easy |
| Water chamber | Frequent wear item | $ | Easy |
| Auto-feed / float valve assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio