Fanem · Model FOG 2140A
Microprocessor-based controls Adjusts temperature administered to patients (servo control) Indicates temperatures at the chamber outlet and close to the patient, as well as humidity supplied to the patient, date and time and length of use by the patient Audible and visual alarms Key for selecting mode of use: Invasive or Noninvasive Key for silencing alarms Enables circuit with heating Double sensor for monitoring temperatures close to the patient and at the water chamber outlet LCD color screen Invasive and Noninvasive mode Compatible for use with mechanical ventilators.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Dry gas to the patient, low or no mist, chamber not warming.
Root cause: Heater element or sensor fault, chamber not seated, or a water-level / feed fault.
Symptoms: High-temperature alarm, heater cuts out.
Root cause: Sensor or control fault, or a chamber run dry.
Symptoms: Chamber runs dry or overfills, water-out alarm.
Root cause: Float valve or auto-feed fault, or a cracked chamber.
Symptoms: Rain-out in the tube, heater-wire alarm.
Root cause: Open heated breathing circuit or a heater-wire adapter fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater base plate / element | Major service part | $$ | Difficult |
| Temperature sensor / probe set | Frequent wear item | $$ | Fairly easy |
| Water chamber | Frequent wear item | $ | Easy |
| Auto-feed / float valve | Periodic replacement | $$ | Moderate |
| Heater-wire adapter | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.