Integrated temperature probe Wall-heated, all-in-one breathing circuits Adjustable temperature and humidity settings Pre-assembled and ready for use.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Dry gas to the patient, low airway temperature, chamber not warming.
Root cause: Heater base plate fault, chamber not seated, a temperature-probe fault, or a water-feed issue.
Symptoms: Rain-out in the tube, heater-wire fault alarm.
Root cause: Open heated breathing circuit or a heater-wire adapter fault.
Symptoms: Implausible temperature, probe-fault alarm, unit will not run.
Root cause: Damaged airway or chamber temperature probe or its connector.
Symptoms: Chamber runs dry or overfills, water-out alarm.
Root cause: Float valve or auto-feed fault, or a cracked chamber.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater base plate | Major service part | $$$ | Difficult |
| Temperature probe set (airway + chamber) | Frequent wear item | $$ | Easy |
| Heater-wire adapter | Periodic replacement | $ | Fairly easy |
| Water chamber | Frequent wear item | $ | Easy |
| Auto-feed / float valve assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.