Teleflex Medical · Model ConcahTherm Neptune
Can be safely used across the continuum of care, from neonates to adults. Uses the safe, convenient ConchaColumn™ and prefilled sterile water reservoir system. Compatible with Hudson RCI® heated-wire breathing circuits. Adjustable temperature and temperature gradients to control circuit condensation. Designed for invasive and noninvasive applications, including noninvasive ventilation, Heated Humidified High Flow Nasal Cannula Therapy (HH-HFNCT) and Continuous Positive Airway Pressure (CPAP). Intuitive user interface. High ventilator output mode compensates for high outlet temperatures and challenging modes of ventilation. Displays and regulates to the desired patient airway temperature.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Dry gas to the patient, low airway temperature, chamber not warming.
Root cause: Heater base plate fault, chamber not seated, a temperature-probe fault, or a water-feed issue.
Symptoms: Rain-out in the tube, heater-wire fault alarm.
Root cause: Open heated breathing circuit or a heater-wire adapter fault.
Symptoms: Implausible temperature, probe-fault alarm, unit will not run.
Root cause: Damaged airway or chamber temperature probe or its connector.
Symptoms: Chamber runs dry or overfills, water-out alarm.
Root cause: Float valve or auto-feed fault, or a cracked chamber.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater base plate | Major service part | $$$ | Difficult |
| Temperature probe set (airway + chamber) | Frequent wear item | $$ | Easy |
| Heater-wire adapter | Periodic replacement | $ | Fairly easy |
| Water chamber | Frequent wear item | $ | Easy |
| Auto-feed / float valve assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Customer complaints were received of power to the unit ceasing and the display going blank.
Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio