Awareness Technology · Model Stat Fax 3300
• Flip-up display screen. • Large 240 x 128 pixel graphical display. • Plug in standard external keyboard for easier data entry (sold separately). • Use 40-column on-board thermal printer, external printer, or PC. • Create a work list, run tests, print patient reports. Stores patient data. • Shows kinetic plots in real time. • Stores control values and creates Levey-Jennings plots for QC. • Optical system is temperature controlled to 37C. • Bichromatic w/ 6 filters and spaces for 2 optional filters (340-700nm). • Adapts to use 12 mm round tubes, 1 cm square cuvettes, or built in flow cell. • Vacuum system speeds sampling and reduces carryover. • Calibration traceable to NIST standard. • Stat Fax readers use IAD filters.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.