Abbott · Model Architect Ci8200
Architect ci8200 from Abbott Diagnostics is an integrated system consolidating clinical chemistry assay and immunoassays on one platform Throughput of up to 1,200 clinical chemistry and 200 immunoassay tests per hour enhances workflow One operator interface for both chemistry and immunoassay testing reduces staff training and delivers one combined report One LIS interface reduces connectivity costs Large onboard reagent capacity of 25 immunoassays and 68 chemistries reduces operator intervention and minimizes sample splitting Robotic Sample Handler (RSH) maximizes analytical productivity with “real time” STAT, automated rerun, and reflex testing capabilities Sample loading of up to 367 samples via priority (35), routine (300), and carousel (32) areas provides continuous sample access Calibrator and quality control materials can be run via the RSH or sample carousel, providing additional loading flexibility Clot detection utilizing pressure differential technology ensures accurate sampling Minimal sample carryover of less than 0.1 ppm increases confidence in results Multiple reagent kit and cartridge sizes allow customization and minimize reagent waste 2-D barcoded reagents provide data on lot numbers, calibration curves, and expiration dates for automatic tracking Extensive calibration and reagent stability reduce reagent usage and operator hands-on time ci-Series 5-position carrier and tray processes multiple sample types, calibration, and quality control material to reduce sample sorting Intuitive indicator lights assist the operator in assessing sample status at a glance Large onboard consumable and buffer supplies provide 5-hour walkaway time.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Controls out of range, calibration fails, flagged results, drift over a shift.
Root cause: Reagent or cuvette issue, lamp ageing, a probe / mixer / wash-station fault, or ISE-module problems.
Symptoms: Short aspiration, no dispense, clot detection, probe-position error.
Root cause: Clogged or bent probe, or a clot in the sample.
Symptoms: High cuvette blanks, non-uniform results by position, rotor-drive error.
Root cause: Worn / scratched cuvettes or a wash fault, or a rotor-drive fault.
Symptoms: Electrolyte results biased or noisy, ISE calibration fails.
Root cause: Aged ISE electrodes / reference, or a pinch-valve / tubing fault.
Symptoms: Multiple station alarms, priming failures, poor wash.
Root cause: Water / DI supply quality, a pump, or a vacuum / pressure regulator.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample / reagent probes | Periodic replacement | $$ | Moderate |
| Photometer lamp | Frequent wear item | $$ | Moderate |
| Reaction cuvettes / segment | Frequent wear item | $$ | Easy |
| ISE electrodes (Na / K / Cl) & reference | Frequent wear item | $$ | Fairly easy |
| Peristaltic / pinch-valve tubing | Frequent wear item | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.