Abaxis · Model Vetscan HM2
Small sample size requirement: consumes approximately25 µl of EDTA-preserved whole blood per CBC measurement. Rapid test turnaround: only 2–3 minutes to results. Advanced, integrated self-cleaning system for optimized performance with very minimal maintenance - expanded automatic aperture cleaning after every run. Simple, intuitive, and easy-to-use software interface makes the analyzer very user friendly, and requires only a short learning curve. Accurate results for out-of-range values: automatic calculations are provided for external 1:5 dilutions in cases of extremely high values or very small sample volumes. Simple, flexible database with large storage capacity: Automatically saves up to 1,000 test records. Data can be downloaded to a USB flash drive or a compatible data-management system. Quality control (QC) results (including Levy-Jennings charts) are stored in the instrument’s QC database for viewing and management. User reminders for replacing the reagent pack and performing simple maintenance task Integrated laboratory results: the HM2 canbe connected to a VetScan Chemistry Analyzer for integrated results reporting. Advanced user features: Multi user mode, customizable normal range, and adjustable analizing strength.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, clog alarm, poor precision.
Root cause: Debris or a clot partially blocking the counting aperture.
Symptoms: WBC or HGB results shifted, high background, reagent-empty or pressure alarms.
Root cause: Reagent pack depleted or expired, or a tubing / pinch-valve fault.
Symptoms: Short samples, no aspiration, probe-position errors.
Root cause: Worn or bent sample probe, or a clot in the sample.
Symptoms: Haemoglobin results biased, blank absorbance off.
Root cause: Photometer lamp aging or a dirty flow cell.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe | Periodic replacement | $$ | Moderate |
| Peristaltic / pinch-valve tubing | Frequent wear item | $ | Moderate |
| Reagent pack | Consumable | $$ | Easy |
| HGB photometer lamp / flow cell | Periodic replacement | $$ | Moderate |
| Aperture assembly | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.