Vital Scientific · Model Selectra Junior
A Random Access, Fully-Automated, "Walk Away" Clinical Chemistry Analyzer. The Vitalab Selectra Junior brings the excellent reputation of the Selectra family into every laboratory. From now on, a small clinical chemistry analyser no longer means that you have to sacrifice quality and performance. The Selectra Junior is the ideal combination in terms of proven technology, performance, versatility and price. It allows you to fulfill demanding requirements in small laboratory settings in the most convenient and cost effective way. The Selectra Junior is a truly open system that can operate using all chemistries that have already been adapted for the Vitalab Selectra. Clinical Chemistry, Special Proteins, Drugs of Abuse, Therapeutic Drugs, Electrolytes - you name it, the Vitalab Selectra Junior does the job without compromising quality. Having a throughput in the range of 100 tests per hour, the Vitalab Selectra Junior is the ideal benchtop random access analyser for small clinics, as a back-up system for medium size laboratories or as a special test analyser.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Controls out of range, calibration fails, flagged results, drift over a shift.
Root cause: Reagent or cuvette issue, lamp ageing, a probe / mixer / wash-station fault, or ISE-module problems.
Symptoms: Short aspiration, no dispense, clot detection, probe-position error.
Root cause: Clogged or bent probe, or a clot in the sample.
Symptoms: High cuvette blanks, non-uniform results by position, rotor-drive error.
Root cause: Worn / scratched cuvettes or a wash fault, or a rotor-drive fault.
Symptoms: Electrolyte results biased or noisy, ISE calibration fails.
Root cause: Aged ISE electrodes / reference, or a pinch-valve / tubing fault.
Symptoms: Multiple station alarms, priming failures, poor wash.
Root cause: Water / DI supply quality, a pump, or a vacuum / pressure regulator.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample / reagent probes | Periodic replacement | $$ | Moderate |
| Photometer lamp | Frequent wear item | $$ | Moderate |
| Reaction cuvettes / segment | Frequent wear item | $$ | Easy |
| ISE electrodes (Na / K / Cl) & reference | Frequent wear item | $$ | Fairly easy |
| Peristaltic / pinch-valve tubing | Frequent wear item | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.