Fast - With unprecedented 18 mm spot size and repetition rates up to 2 Hz, the GentleYAG offers fast treatment with fewer laser pulses. Powerful - 26,333 watts of peak power and capable of delivering fluences up to 600 J/cm2 for vascular applications, and hair removal energies in pulse durations as short as 3ms for unparalleled treatment efficacies, especially finer hairs. Versatile - Non-ablative treatment, including fine lines and wrinkles, and for pseudofolliculitis barbae (PFB), or beard bumps, leg and facial veins, and skin tightening. Treats All Skin Types - Deep penetration with minimal scattering of laser energy makes the GentleYAG ideal for all skin types, especially dark and tanned skin. Multiple Configurations - Choose the GentleYAG that is right for you. GentleYAG lasers come in multiple configurations to fit your specific needs – GentleYAG HR for unparalleled hair removal; GentleYAG SRT for skin rejuvenation and tightening through wrinkle reduction; or the GentleYAG VR for superior leg and facial veins removal. All models are upgradeable.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Weak or no output at the delivery device, treatment ineffective, power not reaching the set value.
Root cause: Laser module / lamp / diode-bar ageing, a damaged delivery fibre / handpiece / applicator, or a driver / power-supply fault.
Symptoms: Unit shuts down after a period, over-temperature or low-flow alarm.
Root cause: Chiller / TEC fault, low or contaminated coolant, a failed pump, or blocked airflow.
Symptoms: Burn marks at the tip, high loss, intermittent output, applicator not recognised.
Root cause: Damaged or contaminated fibre / applicator / connector.
Symptoms: Skin cooling not delivered, cooling / cryogen alarm.
Root cause: Contact-cooling / cryogen supply, a valve, or the tip.
Symptoms: Laser will not fire, key / interlock / footswitch error.
Root cause: Footswitch, door interlock, or key-switch fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Power re-calibration required after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Laser module / lamp / diode bar Power re-calibration required after replacement. | Periodic replacement | $$$$ | Expert-level |
| Delivery fibre / handpiece / applicator | Frequent wear item | $$$ | Fairly easy |
| Chiller / TEC components & coolant filter | Periodic replacement | $$$ | Difficult |
| Contact-cooling tip / cryogen valve (aesthetic) | Periodic replacement | $$ | Moderate |
| Footswitch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio