Stille-Sonesta · Model imagiQ
Lateral tilt around a virtual axis reduces the need for C-arm repositioning Simultaneous free-float and lateral tilt for unparalleled patient access and visualization Quick - Trend function safely & quickly positions the patient in a Trendelenburg position Ultra thin carbon fiber tabletop for optimum imaging at minimum radiation doses Maximized imaging field permits imaging of entire patient anatomy Compact design and small footprint enhance C-arm and clinician access
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Table won't move to commanded position, or moves erratically.
Root cause: Hydraulic pump or electric actuator wear/failure.
Symptoms: Unresponsive or intermittent hand/foot control.
Root cause: Pendant cable, contacts, or wireless receiver wear.
Symptoms: Back, leg, or Trendelenburg section won't lock or articulate.
Root cause: Articulation mechanism or locking pin wear.
Symptoms: Wobble, drift, or difficulty repositioning the table.
Root cause: Caster or locking-wheel mechanism wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Hydraulic pump / electric actuator | Major service part | $$$$ | Difficult |
| Control pendant / remote assembly | Periodic replacement | $$ | Fairly easy |
| Section articulation mechanism | Major service part | $$$ | Moderate |
| Casters / locking wheel set | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.