Mizuho OSI · Model ProAxis 6988
Prone, supine and lateral patient positioning allowing for flexion and extension Radiolucent hinge allows for unrestricted O-arm® and C-arm access Direct visualization of intra-operative changes in spinal alignment Total spine access: cervical, thoracic and lumbar Torso Trolley® integrates patient movement with hinge movement maintaining neutral upper body position with coordinated patient translation Adjustable Chest Slide for optimum patient positioning The IntelliPENDANT®, a software-driven hand pendant, controls all table functions: Floor Locks, Height, Lateral Roll, Trendelenburg/reverse Trendelenburg and Return to Level Removable two-piece carbon fiber supine tops with pad and built-in translation slide convert open frame to a platform for supine and lateral positioning Cervical Mounting Device permits cranial stabilization option ProneView® Protective Helmet System offers head and eye protection for the prone patient Tempur-Pedic® Medical Pad technology incorporated into all patient surfaces. Auxiliary Panel provides a secondary source for controlling table motion Powered Floor Locks.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: Table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Symptoms: Segments loose, accessory rails will not clamp, pads torn, radiolucent top damaged.
Root cause: Worn segment latches, rail clamps, or upholstery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top segment pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio