Berchtold · Model OPERON D 820
1000 lbs of lift through full range of articulation including table top slide Superior positioning of bariatric patients with consistent weight capacity Range of motion from 25 to 42.3 Accommodates standing or seated position for the surgeon Nearly 17 of longitudinal slide Maximum tilt of +/- 20º and Trendelenburg/Rev of 30 º.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: Table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Symptoms: Segments loose, accessory rails will not clamp, pads torn, radiolucent top damaged.
Root cause: Worn segment latches, rail clamps, or upholstery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top segment pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The Surgical Table Cushion (HybriGel Pads) could potentially contain staples that's secure two underlying fabrics, which would be captured on certain scans on patients, such as x-ray.
Table was not lowering.
Difficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.