Berchtold · Model OPERON D 820
1000 lbs of lift through full range of articulation including table top slide Superior positioning of bariatric patients with consistent weight capacity Range of motion from 25 to 42.3 Accommodates standing or seated position for the surgeon Nearly 17 of longitudinal slide Maximum tilt of +/- 20º and Trendelenburg/Rev of 30 º
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Table won't move to commanded position, or moves erratically.
Root cause: Hydraulic pump or electric actuator wear/failure.
Symptoms: Unresponsive or intermittent hand/foot control.
Root cause: Pendant cable, contacts, or wireless receiver wear.
Symptoms: Back, leg, or Trendelenburg section won't lock or articulate.
Root cause: Articulation mechanism or locking pin wear.
Symptoms: Wobble, drift, or difficulty repositioning the table.
Root cause: Caster or locking-wheel mechanism wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Hydraulic pump / electric actuator | Major service part | $$$$ | Difficult |
| Control pendant / remote assembly | Periodic replacement | $$ | Fairly easy |
| Section articulation mechanism | Major service part | $$$ | Moderate |
| Casters / locking wheel set | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The Surgical Table Cushion (HybriGel Pads) could potentially contain staples that's secure two underlying fabrics, which would be captured on certain scans on patients, such as x-ray.
Table was not lowering.
Difficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →