Sorin Group · Model 3T Heater-Cooler System
Extremely short pre-cooling times due to the small volume of the tank Separate tanks enable switching between warm and cold cardioplegia without delay Cooling with the ice container is not necessary during routine operation due to the strong performance of the compressor assembly An independent safety system prevents the water temperature from reaching critical values The two-chamber pumps minimize any pressure inside the heat exchanger The suction stage of the pumps enables a smooth emptying of the heat exchanger and the tubing
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
Common R&R.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
| Blower fan | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
Potential colonization of organisms, including Mycobacteria, in Sorin Heater Cooler Devices, if proper disinfection and maintenance is not performed per Instructions for Use.
On August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →