Medivance · Model Arctic Sun 5000
Fast initiation of treatment Simple programming capabilities Comprehensive built-in training module Easy access to treatment data Precise Temperature Control: Enhanced treatment algorithm over the Arctic Sun 2000® Temperature Management System Rapid Clinician Training: Intelligent automation and built-in education tutorial Comprehensive Implementation Program: Arctic Circle™ Clinical Resource Program.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Servo control cannot achieve or hold the target core temperature, slow rate of change, water-temperature limit reached.
Root cause: ArcticGel pad contact / coverage, a circuit air lock, low water level, a pump fault, or a heater / chiller (TEC) fault - use the Diagnostic screen (flow, pressure, water-sensor readings).
Symptoms: Circuit alarm, delivery pauses, flow low on the Diagnostic screen.
Root cause: Kinked or clamped pad lines, a disconnected pad, an air lock, or a pump / manifold fault.
Symptoms: Core / surface temperature reads implausibly, probe-fault alarm, servo unavailable.
Root cause: Damaged temperature probe or its adapter cable.
Symptoms: Low-water alarm, water around the unit, will not fill.
Root cause: Reservoir leak, a fill-port issue, or worn internal tubing / seals.
Symptoms: Water or patient temperature disagrees with a reference on the calibration check.
Root cause: Temperature-sensor or control calibration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| ArcticGel pads / hoses | Consumable | $$ | Easy |
| Patient temperature probe / adapter cable | Frequent wear item | $$ | Easy |
| Circulation pump | Periodic replacement | $$$ | Difficult |
| Heater / chiller (TEC) module | Major service part | $$$$ | Difficult |
| Water reservoir / internal tubing kit | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.
Potential for an inaccuracy of the primary patient temperature reading that is used to monitor and control patient temperature.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.