Baxter International · Model RK-2000
The Baxter RK 2000 hypothermia unit supplies temperature-controlled water through a connector hose to a blanket in order to regulate the temperature of hypothermia and hyperthermia patients. The unit is able to control the water temperature by mixing hot and cold water in solenoid valves underneath the microcontroller control. A probe senses the patients temperature and fluid and displays it on the LED control panel for for ease of patient management. Additionally, this unit has three different modes; manual mode which allows the user to set the blanket temperature and the unit will bring the blanket to that desired set point, auto mode which allows the user to set the patient temperature and the unit will adjust the blanket temperature to achieve the desired patient temperature, and monitor mode in which the user can simply monitor the patients temperature without providing any therapy. The blankets can be used either on top or underneath the patient and a second blanket can be attached to this unit for more rapid therapy.This unit also has four casters attached to it allowing it to be rolled around.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
Common R&R.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Blower fan | Periodic replacement | $$ | Fairly easy |
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
There is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.