Skytron · Model 6702 Hercules
210° top rotation Removable back and leg sections simplify accessory setup Power beach chair positioner allows for automatic, one-touch setup of shoulder procedures Table height adjustable from 23-5/8” to 41-1/2” Low table height allows surgical team to comfortably operate at waist height on bariatric patients Slim-profile based will not conflict with imaging equipment.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand-control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: The table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.