Skytron · Model 6000
• Control from any point around the table with one-touch pendant control. • Positioning is safe and easy using the table’s easy-to-read symbols. • Provides adequate toe space and mayo stand clearance with its narrow base design. • Features a radiolucent top that provides clear, sharp patient imaging capability. • Easily clean the durable, stainless steel base and column. Also easily clean between cases using swivel casters, which provide side movement. • Fully mechanical manual brake release utilizes no battery power for emergency use. • Provides C-arm access to nearly all of a patient’s anatomy. • Features a reversible head section you can adjust to accommodate tall patients. You can also use it as a foot extension during patient transfer.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Table won't move to commanded position, or moves erratically.
Root cause: Hydraulic pump or electric actuator wear/failure.
Symptoms: Unresponsive or intermittent hand/foot control.
Root cause: Pendant cable, contacts, or wireless receiver wear.
Symptoms: Back, leg, or Trendelenburg section won't lock or articulate.
Root cause: Articulation mechanism or locking pin wear.
Symptoms: Wobble, drift, or difficulty repositioning the table.
Root cause: Caster or locking-wheel mechanism wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Casters / locking wheel set | Major service part | $$ | Fairly easy |
| Hydraulic pump / electric actuator | Major service part | $$$$ | Difficult |
| Control pendant / remote assembly | Periodic replacement | $$ | Fairly easy |
| Section articulation mechanism | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.