Siemens Healthineers · Model ADVIA 2400
Up to 2400 tests/hour (1800 photometric and 600 ISE) Rapid, consistent 2-second cycle time Top-of-class throughput of 200 tubes per hour when running ISEs and up to 9 photometric tests Automatic Sample Retain technology, auto-repeat, auto-dilution and auto-reflex testing Point-in-space aspiration offers connectivity to track or rack handler without robotics or special hardware Redundant sample loading for built-in backup and easy handling of non-routine samples Direct sampling from track prevents detaining tubes at analyzer and reduces aliquotint Minimal-maintenance ISE No-maintenance 37oC incubation fluorocarbon bath oil Auto-calibration and auto-QC from refrigerated on board calibrator and controls Microvolume technology with 32,000 tests onboard and up to 60 days onboard stability maximizes walkaway time Sample integrity check for icterus, hemolysis and lipemia Sample probe with automatic clot/clog detection, liquid level sensing, short sample detection and crash protection Advances in reagent onboard stability, calibration frequency, interference reduction and linearity expansion.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Controls out of range, calibration fails, flagged results, drift over a shift.
Root cause: Reagent or cuvette issue, lamp aging, a probe / mixer / wash-station fault, or ISE-module problems (electrodes / reference).
Symptoms: Short aspiration, no dispense, clot detection, probe-position error.
Root cause: Clogged or bent probe, or a clot in the sample.
Symptoms: High cuvette blanks, non-uniform results by cuvette position, tray-drive error.
Root cause: Worn / scratched cuvettes (or a wash-fault on reusable cuvettes) or a tray-drive fault.
Symptoms: Electrolyte results biased or noisy, ISE calibration fails.
Root cause: Aged ISE electrodes or reference, or a pinch-valve / tubing fault.
Symptoms: Multiple station alarms, priming failures, poor wash.
Root cause: Water-supply / DI quality, a pump, or a vacuum / pressure regulator.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample / reagent probes | Periodic replacement | $$ | Moderate |
| Photometer lamp | Frequent wear item | $$ | Moderate |
| Reaction cuvettes | Frequent wear item | $$ | Easy |
| ISE electrodes (Na / K / Cl) & reference | Frequent wear item | $$ | Fairly easy |
| Peristaltic / pinch-valve tubing | Frequent wear item | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.