Siemens Healthineers · Model ADVIA 2120i
Save laboratory personnel’s time, improve productivity and reduce costs with practical automation. Results that are right the first time – Quality results for WBC differential, RBC, reticulocyte, PLT and CSF testing. Scalable solution: From standalone to complete automation Save laboratory personnel’s time Only the ADVIA 2120i system gives you Practical Automation - the unique ability to eliminate the majority of manual steps commonly performed for maximum speed and productivity. Proven technology, reference methods, and onboard delta checks reduce manual slide reviews. Fast, convenient, rack-based sampling: Any tube. Any time. High throughput of up to 120 samples per hour. Results that are right the first time. Efficient auto-validation procedures Maximize productivity Scalable solutions Expanding your capabilities is easier and more affordable than ever with the ADVIA 2120i system. With integrated and automated solutions from Siemens, you can scale from stand-alone instrument to complete automation as your productivity needs change. Optional ADVIA® Autoslide Slide Maker Stainer gives you up to 120 high-quality slides per hour to scale up your productivity. Network connectivity via ADVIA CentraLink® Networking Solution allows data management consolidation and QC management from any client workstation. From truly practical automation to complete automation with the same instrument Ease of use The ADVIA 2120i system simplifies and streamlines all aspects of testing, allowing you to work faster and more efficiently so you can focus on other critical tasks. Touch-screen monitor. Unique reagent Time Pacs. Three modes of aspiration: autosampler, open mode, and unique manual closed-tube sampling. Unique Unifluidics technology - reduced fluidics, no pinch valves, automated daily cleaning. Completely automated start-up Quality results - Results that are right the first time The ADVIA 2120i system uses gold-standard peroxidase staining for WBC differential testing. Fewer reruns. Lowest overall review rate. Faster turnaround time. Lowest linearity - 0.02-400k The system also offers state-of-the-art technology for RBC, reticulocyte, PLT and CSF testing. RBC. True RBC differential. Consistent morphology. Reduced manuals. Unique dual-hemoglobin measurement capability - eliminates interference from lipemia Automated Reticulocyte Analysis. Direct measurement, not calculated. Excellent in pediatric and dialysis populations PLT. Two dimensions - size and refractive index. Measures up to 60 fl. Includes large platelets in count. Excludes fragments and microcytic RBCs CSF. Results in <5 minutes. Linear to zero with three-part differential. Automated counting provides quality control for consistency.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.