Tattoo Removal. Pigmentation Lesions - When treating benign pigmented lesions and tattoo ink, the safety profile of the Tri-Lase nanosecond pulse width makes it possible to attain reliable and predictable clinical results with minimal risk of adverse reactions and textural changes to the skin. The PhotoAcoustic power of the Tri-Lase selectively shatters and destroys the target chromophore while preserving the surrounding healthy cellular structures. The macrophages of the skin’s immune system will then start cleansing the fragments from the body resulting in impressive clinical results that are easily obtained and repeatable. Atrophic Acne Scars. Photodamaged Skin Lesions. Mild to Moderate Facial Wrinkles. Uneven Pigment Seborrhic Keratosis - Sandstone’s engineering has implemented a unique extended ablation technology that amply improves patient comfort and minimizes downtime. By combining a proprietary fluence level with an extra-long pulse width, it precisely ablates the skin without exposing the underlying dermis. Thus, the gentle nature of the laser allows you to safely and effectively treat highly sensitive skin regions on the neck, back of the hands and around the peri-oral and peri-orbital areas without anesthesia.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Weak or no output at the delivery device, treatment ineffective, power not reaching the set value.
Root cause: Laser module / lamp / diode-bar ageing, a damaged delivery fibre / handpiece / applicator, or a driver / power-supply fault.
Symptoms: Unit shuts down after a period, over-temperature or low-flow alarm.
Root cause: Chiller / TEC fault, low or contaminated coolant, a failed pump, or blocked airflow.
Symptoms: Burn marks at the tip, high loss, intermittent output, applicator not recognised.
Root cause: Damaged or contaminated fibre / applicator / connector.
Symptoms: Skin cooling not delivered, cooling / cryogen alarm.
Root cause: Contact-cooling / cryogen supply, a valve, or the tip.
Symptoms: Laser will not fire, key / interlock / footswitch error.
Root cause: Footswitch, door interlock, or key-switch fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Power re-calibration required after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Laser module / lamp / diode bar Power re-calibration required after replacement. | Periodic replacement | $$$$ | Expert-level |
| Delivery fibre / handpiece / applicator | Frequent wear item | $$$ | Fairly easy |
| Chiller / TEC components & coolant filter | Periodic replacement | $$$ | Difficult |
| Contact-cooling tip / cryogen valve (aesthetic) | Periodic replacement | $$ | Moderate |
| Footswitch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio