Roche · Model Reflotron Plus
Quick Results. Immediate ready for measurement standby mode. 30 micro liter whole blood, plasma or serum. Quantitative results after 2-3 minutes. Easy to use. Navigation via display. No reagents preparing, No calibration necessary. Automatic parameter detection. Economic testing. Compact bench top analyzer. Single test determination. Long stability of the reagents.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Result does not fit the clinical picture, on-board or liquid QC fails.
Root cause: Expired or heat-damaged reagent strips, a wrong-volume / clotted / haemolysed sample, or use outside the operating temperature range.
Symptoms: Repeated optics or reflectance errors, strip not read correctly.
Root cause: Dirty or aged optics / lamp, or a contaminated measurement chamber.
Symptoms: Strip not transported or the magnetic code not read.
Root cause: Worn strip-transport mechanism or a magnetic-head fault.
Symptoms: Measurement-chamber temperature not at 37 C, results drift.
Root cause: Heater or temperature-sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Refrigerate / store per the insert; heat and expiry degrade them.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reagent strips Refrigerate / store per the insert; heat and expiry degrade them. | Frequent wear item | $$ | Easy |
| Optics / lamp | Periodic replacement | $$ | Moderate |
| Strip-transport / magnetic-head assembly | Major service part | $$ | Moderate |
| Heater / temperature sensor | Major service part | $$ | Moderate |
| Liquid QC / Precinorm | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Coating defect may result is falsely decreased patient and/or control results.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.