Roche · Model Cobas 8000
• Screen sharing with the analyzer software and cobas e-services on a single screen for convenient instrument monitoring. • Traceability information for easy tracking of calibration and reagent information offers more transparency. • User defined fully automated and selective rerun and reflex testing. • Connectivity of up to 3 cobas 8000 modular analyzer series in a single cobas 8000 data manager. • Technical validation of sample result in combination with rule programming improves and accelerates the result release process. • QC management of all connected instruments on a single screen supports flexible integration of local requirements for QC standards and rules (e.g. Levey-Jennings plots and diagrams, Westgard rules,…etc.). • High speed • Intelligent sample dynamics • Intelligent workflow management • Premium quality & safety • Multidimensional modularity
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Low signal, wavelength drift.
Root cause: Lamp/LED or monochromator aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Frequent R&R on high-throughput analyzers.
Consumable - replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
| Sample aspiration probe Frequent R&R on high-throughput analyzers. | Periodic replacement | $$$ | Moderate |
| Peristaltic pump tubing Consumable - replace proactively. | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.
AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
A software failure may incorrectly set the system settings to "default" settings, creating a risk of incorrect results.
Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended. This may lead to carryover and medical risk cannot be excluded.
When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →