Extracted directly from this unit's own service manual.
1An execution error in system processing.
Troubleshoot Executive Processor Board and Video Generator Board.
2Data recorder - Log only fault.
Replace the Executive Processor Board.
3An error occurred reading the internal mass storage device.
Replace the Executive Processor Board.
4An error occurred while saving data to the internal mass storage device.
Replace the Executive Processor Board.
5Cannot write data onto the internal mass storage device.
Replace the Executive Processor Board.
6A non-isolation Digital Signal Processor (DSP) error detected by software.
Replace the Front End Board.
7An isolation Digital Signal Processor (DSP) error detected by software.
Replace the Front End Board.
8An error with a connected CARDIOSAVE Trainer. If no Trainer is present, this is an error with the Front End Board. If a trainer was connected, replace the Trainer. If no Trainer was present, replace the Front End Board.
10A software error was detected during pressure source selection.
No service action required.
11A software error was detected during ECG source selection.
No service action required.
12A software error was detected during fiber-optic sensor calibration.
No service action required.
13A software error was detected during OPERATION MODE selection.
No service action required.
14A software error was detected during timing selection.
No service action required.
15An error was detected during Auto OPERATION MODE Assisted Beat Event Detection (ABED)
No service action required.
16The software processed an invalid trigger request. Verify proper operation of the pressure channel. This may also be caused by a kinked IAB, bad pressure transducer, or pumping with the trainer on no augmentation.
Verify proper operation of the pressure channel.
17A software error was detected during deflation surveillance.
No service action required.
18A software error was detected during inflation surveillance.
No service action required.
19A software error was detected during interval prediction surveillance.
No service action required.
20A software error was detected during pattern identification package.
No service action required.
21A software error was detected during slider package.
No service action required.
22A software error was detected during timing calculation.
No service action required.
23A software error was detected during timing control.
No service action required.
24A software error was detected during trigger change.
No service action required.
25A software error was detected during Unassisted Beat Event Detection (UBED).
No service action required.
26A software error was detected during timing support.
No service action required.
27A software error was detected during triggering.
No service action required.
28A software error was detected during ECG trigger.
No service action required.
29A software error was detected during BP trigger.
No service action required.
30A software error was detected during internal trigger.
No service action required.
31A software error was detected during Pacer V/AV trigger.
No service action required.
32A software error was detected during the Trigger Status State Machine (TSSM) package.
No service action required.
33A software error was detected in the Timing Trigger and Source (TTS) manager package.
No service action required.
34A software error was detected in the utilities package.
No service action required.
35A Head to Base connection error was detected while in System Diagnostics. The fault may be caused by missing data or a bad packet. If fault persists, troubleshoot Base to Head connections including the coiled cord and Video Generator Board.
If fault persists, troubleshoot Base to Head connections including the coiled cord and Video Generator Board.
36A critical software failure generated while in System Diagnostics.
Restart the unit.
37Autofill Failure. Refer to table in section 2.3.3.1 on page 54 for Code 37 - Autofill Failure Troubleshooting and Comment Strings.
Refer to table in section 2.3.3.1 on page 54 for Code 37 - Autofill Failure Troubleshooting and Comment Strings.
38A leak was detected during the Autofill process while catheter alarms were disabled. This is a log only fault.
Troubleshoot the system for leaks. Evaluate the transducers. If the fault persists, replace the Front End board.
39Blood presence was suspected during the Autofill process while catheter alarms were disabled. This is a log only fault.
Troubleshoot the system for leaks. Evaluate the transducers. If the fault persists, replace the Front End board.
40The print buffer queue is full. The printer is not processing data correctly and cannot print.
Restart the system. Check printer connections and replace the printer and/or associated connections. If fault persists, replace the Executive Processor Board.
42The printer buffer stack (memory pool) is full.
Restart the system. If fault persists, replace the Executive Processor Board.
43The printer failed to acquire a buffer stack destination.
Restart the system. If fault persists, replace the Executive Processor Board.
44The printer circular queue is full.
Restart the system. If fault persists, replace the Executive Processor Board.
45A memory leak in the printer service software.
Restart the system. If fault persists, replace the Executive Processor Board.
46A printer message failed to send to a destination queue.
Restart the system. If fault repeats, replace the Executive Processor Board.
47A failure to read data from the printer to the Executive Processor Board.
Restart the system. If the error repeats after restarting, check printer connections and replace printer assembly and/or cables. If fault persists, replace the Executive Processor Board.
48A failure in the printer start timer of the software operating system.
Replace the Executive Processor Board.
49A failure in the printer create message queue of the software operating system.
Replace the Executive Processor Board.
50Repair Information section header.
51Repair Information section header.
115Kernel application program interface error. (Software Critical Error) Examine the fault code Address Space. If the Address Space is "UIHeadDis", "UIHeadSup", or "VideoBIT", replace the Video Generator Board. If another address space is listed, replace the Executive Processor Board.
Examine the fault code Address Space. If the Address Space is "UIHeadDis", "UIHeadSup", or "VideoBIT", replace the Video Generator Board. If another address space is listed, replace the Executive Processor Board.
116A hardware and/or software failure that affects base to head communication over the coil cord.
Troubleshoot Base to Head connections including the coiled cord and Executive Processor Board.
117A software detected failure. A critical device resource was unavailable.
Replace the Executive Processor Board. If the fault persists troubleshoot respectively: 1. If the associated string is "Hpi device fault" replace the Front End Board. 2. If the associated string is "Couldn't change register on IODevice", replace the Solenoid Driver Board.
118A critical error detected in the hardware watch dog circuit on the Solenoid Driver Board. The isolation (K4) or inflate (K8) solenoids remained energized for more than 2 seconds or the deflate (K7)