Edwards Lifesciences · Model Aquarius
Low and high volume haemofiltration capabilities. Wide range of Aquamax filters choice to optimize patient benefit - Adult and Pediatric. One pre-connected tubing set for all treatments. Provide step by step, on-screen instructions. Simple set-up with molded tubing set. Automated priming.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: BLD AD/steps too low (<250), zero, or too high (>800) during calibration; false blood-leak alarms.
Root cause: Scratched detection chamber, dirty or wrongly-inserted mirror (silver spot up), LED/sensor misalignment, BLD-module potentiometer out of adjustment, or a failed BLD module / Master CPU.
Symptoms: Flow error on blood, dialysate, filtrate or substitution pump; occlusion or pressure alarm; inaccurate fluid balance.
Root cause: Worn pump rollers/race, a pump-door / occlusion sensor fault, or a motor/encoder fault.
Symptoms: Fluid-balance alarm, implausible filtration or substitution volume, treatment stops on balance error.
Root cause: One of the two weighing scales / load cells out of calibration or mechanically bound.
Symptoms: Access, return, filter or effluent pressure reads wrong; frequent pressure alarms.
Root cause: One of the four pressure transducers drifted or a protector wetted.
Symptoms: Return fluid too cool or warm, temperature alarm, air detected at the degassing chamber.
Root cause: Heater element, temperature sensor at the degassing chamber, or a degassing fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate fluid balance after replacement.
Single-treatment consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump roller / race kit | Periodic replacement | $$$ | Difficult |
| Weighing scale / load cell Recalibrate fluid balance after replacement. | Major service part | $$$ | Difficult |
| Pressure transducer / protector set | Periodic replacement | $$ | Moderate |
| Blood leak detector module (BG-E431) | Major service part | $$$ | Difficult |
| Substitution-fluid heater element | Major service part | $$$ | Difficult |
| Pre-connected tubing set Single-treatment consumable. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.