Eschmann · Model T20 -m+
• T20 - m+ model features larger castors and unique brake pedal design • Full length radio translucent top with built-in x-ray cassette tunnel accessible from either end • 720-1120mm height range • 450kg maximum patient weight • Powered 250mm table top traverse • 600mm wide table top (across sidebars) • Corded handset controlling Trendelenburg, reverse Trendelenburg, lateral tilt, flexion/extension, traverse, auto level, height, and battery life indicators • Offset table-top break and offset slim line column for superior imaging and access • Column covers in stainless steel • Battery powered (mains rechargeable) • Head and leg sections are removable and interchangeable, with an auto-locking mechanism • Covers to the top and base are made of scratch resistant and colourful acrylic capped ABS mouldings to minimise potential sites for cross contamination and provide a wash down surface to the tabletop • Simple mattress fixings, no velcro • Special one-piece mattress covering break position (long and short trunk sections) to combat fluid ingress and to provide superior patient support • Supplied with 50mm deep mattress
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Table won't move to commanded position, or moves erratically.
Root cause: Hydraulic pump or electric actuator wear/failure.
Symptoms: Unresponsive or intermittent hand/foot control.
Root cause: Pendant cable, contacts, or wireless receiver wear.
Symptoms: Back, leg, or Trendelenburg section won't lock or articulate.
Root cause: Articulation mechanism or locking pin wear.
Symptoms: Wobble, drift, or difficulty repositioning the table.
Root cause: Caster or locking-wheel mechanism wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Section articulation mechanism | Major service part | $$$ | Moderate |
| Casters / locking wheel set | Major service part | $$ | Fairly easy |
| Hydraulic pump / electric actuator | Major service part | $$$$ | Difficult |
| Control pendant / remote assembly | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →