Eschmann · Model T20-a : T20-s
Full length radio translucent top with built-in x-ray cassette tunnel accessible from either end. 720-1120mm height range. 450kg maximum patient weight. Powered 250mm table top traverse. 600mm wide table top (across sidebars). Corded handset controlling Trendelenburg, reverse Trendelenburg, lateral tilt, flexion/extension, traverse, auto level, and height, Offset table-top break and offset slimline column for superior imaging and access. Column covers in stainless steel. Battery powered (mains rechargable). Foot pedals on base to control brakes, wheels and castors for stability, straight-line steering and 360 degree mobility. Head and leg sections are removable and interchangeable, with an auto-locking mechanism. Covers to the top and base are made of scratch resistant and colourful acrylic capped ABS mouldings to minimise potential sites for cross contamination and provide a wash down surface to the tabletop. Simple mattress fixings, no velcro. Special one-piece mattress covering break position (long and short trunk sections) to combat fluid ingress and to provide superior patient support.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: Table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Symptoms: Segments loose, accessory rails will not clamp, pads torn, radiolucent top damaged.
Root cause: Worn segment latches, rail clamps, or upholstery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top segment pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio